Flukes 1 kilogram answer to pressure calibration
Of course size matters -and so does weight. Particularly if you have to carry your test tools around with you.
The Fluke 718 pressure calibrators are about a third of the size of comparable instruments and weigh just one kilogram.
With its built-in pressure/vacuum pump, the Fluke 718 offers a complete solution for on-site pressure maintenance and calibration.
Out in the field, youll really notice the difference.
| Measurement Accuracy |
| Pressure |
| Fluke 718 30G |
|
|
|
| Range: |
-850 mbar to 2 bar (internal sensor) Over Pressure 3xFS |
| Resolution: |
0.1 mbar |
| Accuracy: |
0.05% full scale |
| Functions: |
Zero, Min, Max, Hold, Damp |
|
Note: for use with non-corrosive gasses |
|
|
| Fluke 718 100G |
|
|
|
| Range: |
-850 mbar to 7 bar (internal sensor) Over Pressure 3xFS |
| Resolution: |
1 mbar |
| Accuracy: |
0.05% of full scale |
| Functions: |
Zero, Min, Max, Hold, Damp |
|
Note: for use with non-corrosive gasses |
|
|
| Fluke 718 series |
|
|
|
|
via pressure module connector |
| Range: |
29 pressure modules, 2,5 mbar to 700 bar |
|
Over pressure per pressure module specs |
| Resolution: |
Per pressure module specs |
| Accuracy: |
To 0.025% of full span, per pressure module specs |
| Functions: |
Zero, Min, Max, Hold, Damp |
|
Note: media compatibility per pressure module specs, |
|
|
| Source Pressure, built in pump |
|
| Range: |
-820 mbar to full scale |
|
|
|
Supported Pressure Units; psi, in. H20
(4 °C), in. H20 (20 °C), cm H20 (4 °C), |
|
cm H20 (20 °C), bar, mbar, kPa, inHg, mmHg, kg/cm |
|
| Technical Data |
| Loop Power |
| Range: |
24 V dc |
| Accuracy: |
+/- 10% |
|
Note: Drive; 20 mA into 1000 W for battery >6.8V; 700 W for battery 5.8 to 6.8 V |
|
| Measure Accuracy: |
| Range: |
0 to 24 mA |
| Resolution: |
0.001 mA |
| Accuracy: |
0.025% + 1 count |
|
| Display |
|
LCD, 5 digit pressure and current simultaneous |
|
| Environmental Specifications |
| Operating temperature |
|
| Storage temperature |
|
| Humidity (Without Condensation) |
|
95% (10ºC to 30ºC) |
|
75% (30ºC to 40ºC) |
|
45% (40ºC to 50ºC) |
|
35% (50ºC to 55ºC) |
|
| Operating Altitude |
|
| Safety Specifications |
| Safety |
|
CSA C22.2 No. 1010.1:1992 |
|
| Maximum voltage |
|
| Mechanical & General Specifications |
| Size |
|
210 mm L x 87 mm W x 62 mm |
|
216 mm L x 94 mm W x 67 mm (with holster) |
|
| Weight |
|
Approx. 737g |
|
Approx. 992g (with holster) |
|
| Warranty |
|
| Shock & Vibration |
|
Random, 2G, 5-500 Hz |
|
1m drop test |
|
| Battery type |
|
Two 9V batteries, ANSI/NEDA 1604A or IEC 6LR61 |
|
Battery life: 4 to 20 hours, typical, depending on functions |
|
|