Double Your Power.
A maintenance and calibration tool that will be at home in every instrumentation technician's tool box, the Fluke 787 ProcessMeter combines a DMM and a Loop Calibrator in one rugged, handheld tool, for about what you would expect to pay for a loop calibrator alone. Based on the trusted Fluke 87 DMM, the 787 adds the ability to measure, source, and simulate dc loop current with 0.05% accuracy and 1 microamp resolution.
Fluke Process Meters - ProcessMeter Model 787 offers:
• Simultaneous mA and % of scale readout on mA output.
• 25% Manual Step plus Auto Step and Auto Ramp on mA output.
• Clear LCD with backlight; 4,000 counts (30,000 counts for dc current)
• Min/Max/Average/Hold/Relative modes.
• Externally accessible battery for easy battery changes.
• 1000V overload protection on V, ohms, frequency.
• 150V overload protection on mA, backed up by 440 mA 1000V
| Technical Data |
| Current Measurement |
| Range: |
0-1 A |
| Resolution: |
1 mA |
| Accuracy: |
0.2% + 2 LSD |
|
|
| Range: |
0-30 mA |
| Resolution: |
0.001 mA |
| Accuracy: |
0.05% + 2 LSD |
|
| Current Sourcing |
| Range: |
0-20 mA or 4-20 mA |
| Resolution: |
0.05% of span |
| Drive Capability: |
500 W @ 24 mA |
|
| Frequency |
|
to 19.999 kHz, 0.005% + 1 LSD |
|
| Continuity |
|
Beeps for resistance < 100 ohms |
|
| Voltage measurement |
| Range: |
0-1000 V AC or DC |
| Resolution: |
0.1 mV to 1.0 V |
| Accuracy: |
0.1% Rdg + 1 LSD (VDC) |
|
| Resistance measurement |
|
| Diode test |
|
2.4 V shows diode voltage drop |
|
| Environmental Specifications |
| Operating temperature |
|
| Storage temperature |
|
| Humidity (Without Condensation) |
|
95% up to 30 ºC |
|
75% up to 40 ºC |
|
45% up to 50 ºC |
|
35% up to 55 ºC |
|
| Temperature Coefficient |
|
0.05 x (specified accuracy) per ºC (for temperatures < 18 ºC or > 28 ºC) |
|
| Safety Specifications |
| Electrical Safety |
|
Designed in accordance with IEC 1010-1, ANSI/ISA S82.01-1994 and CAN/CSA C22.2 No. 1010.1-92 |
|
| Overvoltage category |
|
| Maximum voltage |
|
| Mechanical & General Specifications |
| Size |
|
| Weight |
|
| Batteries |
|
Single 9V Alkaline battery (ANSI/NEDA 1604A or IEC 6LR61) |
|
| Warranty |
|
| Battery life |
|
50 hours typical (measurement), 12 hours typical (sourcing 12 mA) |
|
| Shock & Vibration |
|
Random, 2g, 5-500 Hz |
|
1 meter drop test |
|
| Display current and % of span |
|
| Auto step, auto ramp |
|
|