Process Control Home
 
Alpha Controls
 

Model 9101 Zero Point

Model 9101 Zero Point

Model,9101,Zero,Point

Features

Bath-quality stability in a portable ice-point reference

Easy re-calibration for long-term reliability

Ready-light frees user’s time and attention

Solid-state cooling technology


Specifications

Temperature Range

0°C (32°F)

Stability

±0.005°C

Total Instrument Error

±0.02°C, typical; ±0.05°C max. (18–25°C ambient)

Stabilization Time

Approx. 30 minutes
(the ready lamp indicates stable control at 0°C)

Temperature Coefficient

±0.005°C/°C

Size

12.25” H x 8.5” W x 5.75” D
(311 x 216 x 146 mm)

Power

115 VAC (±10%), 1 A or 230 VAC (±10%), 0.5 A, specify, 50/60 Hz, 125 W

Well Dimensions

2 wells 0.25" dia. x 6" D (6.4 x 152 mm), 1 well 0.28" dia. x 6" D (7 x 152 mm). Includes one set of telescoping inserts to provide various smaller diameters

Weight

12 lb. (5.4 kg)

NIST-Traceable Calibration

Data at 0°C


Have you been thinking about buying a zero point dry-well? Forget those ugly-looking units the competition makes. Now you can get a great-looking and great-performing zero point dry-well from Hart Scientific.

The Hart Model 9101 has three test wells for inserting more than one probe at a time. All three wells are stable to ±0.01°C. One well accommodates changeable inserts for varying probe diameters.

The Model 9101 takes advantage of the latest solid-state cooling technology rather than relying on older, less reliable sealed-water-cell devices. This eliminates the possibility that the sealed-water cell will freeze and burst while transporting the unit to field locations. And our solid-state cooler is run by an adjustable electronic controller that can be recalibrated in your lab for convenient recertification. Simply place a certified standards thermometer in one of the wells (See Improving Dry-Well Calibrations) and, if needed, tweak the 9101 controller until the standards thermometer reaches equilibrium at 0°C.

Since the unit is completely self-contained and doesn’t require any user settings, you can run it on demand for instant access to an accurate, traceable zero point. Set it up with the reference junction of a thermocouple for high- accuracy thermocouple measurements.

Less costly than refrigerated baths, more accurate and less problematic than ice baths and more durable and better-looking than competitive units using sealed-water cells, the Hart Model 9101 Zero Point Dry-Well is a great choice for any calibration lab.



Powerful Solutions for Instrumentation & Control Problems!

Model 9101 Zero Point

Site Designed By Panther Marketing