Secondary Reference Temperature Standard
Probes -Models 5612, 5613 and 5614

|
Features
|
|
Affordable wide-range accuracy
|
|
Excellent stability
|
|
Reference-grade platinum sensing element
|
|
Specifications
|
|
Resistance
|
Nominal 100W (±0.1W)
|
|
Temperature Coefficient
|
0.003925 ohms/ohm/°C nominal
|
|
Temperature Range
|
200°C to 420°C (transition and cable temperature 150°C maximum)
|
|
Transition Temperature
|
5°C to 200°C
|
|
Drift Rate
|
±0.01°C at 0°C per year maximum, when used periodically to 400°C
|
|
Sheath Material
|
Inconel 600
|
|
Leads
|
Teflon-insulated, silver-plated stranded copper, 22 AWG
|
|
Termination
|
Four gold-plated spade lugs are standard. Other options available. See Ordering Information.
|
|
Hysteresis
|
< 0.01°C at 0°C using 196°C and 420°C as the end points
|
|
Immersion Effects
|
Reading will not vary more than 0.005°C when the probe immersion is varied between 4 inches and 10 inches in an ice bath (5614).
|
|
Calibration
|
Includes NIST-traceable calibration and table with R vs. T values in 1°C increments from 183°C to 500°C. The 5614 and 5612 are calibrated to 420°C and the 5613 to 300°C. ITS-90 coefficients included.
|
|
Probe Accuracy (includes calibration uncertainty and short-term stability)
|
±0.018°C at 196°C
±0.018°C at 0°C
±0.019°C at 200°C
±0.023°C at 420°C
|
|
Time Constant
|
Nine seconds typical for 63.2% response to step change in temperature in water flowing at 3 feet per second
|
|
Size: 5612
5613
5614
|
0.187" dia. x 9"
0.187" dia. x 6"
0.25" dia. x 12"
|
|
High Flexibility Cable Option
|
Standard 5612, 5613, and 5614 PRTs come with a 6-foot white Teflon cable that withstands temperatures from 100°C to 250°C. For super-flexible black PVC cable that can be exposed to temperatures from 0°C to 90°C, add -B to the model number. (A 5614 PRT with spade lug terminations and PVC cable would be ordered as 5614-S-B.) No additional charge is added for PVC cable.
|
Need a durable but accurate sensor for use in the factory, field, or lab? The Model 5614 Secondary Temperature Standard is the answer.
The 5614 is a Platinum Resistance Thermometer (PRT) thats 12 inches long with an Inconel 600 sheath and a 1/4" outside diameter. It is designed to be used as a transfer device from the highest laboratory standards to industrial or second-tier lab locations. It has short-term accuracy of ±0.02°C at 200°C.
The element is constructed of reference-grade platinum wire (99.999% pure) for excellent stability. The wire is wound in a coil and placed in a mandrel where it is uniformly supported in a manner to virtually eliminate hysteresis. The electrical configuration is a four-wire current-potential hookup to eliminate effects of lead-wire resistance.
These Inconel-sheathed probes have a partially supported sensing element, making them more durable than SPRTs. The element is protected in an ultrahigh-purity ceramic case with a hermetic glass seal to improve output stability by locking out moisture and contaminants.
This probe comes calibrated in accordance with ITS-90, which makes it compatible with many excellent readout devices, including Harts 1575 and 1590 Super-Thermometers, 1560 Black Stack, and 1502 Tweener. It bridges the gap between a 100-ohm industrial RTD and an SPRT.
For those needing faster thermal response, or where diameter and immersion depth are problems, order the 6-inch 5613 or the 9-inch 5612. These probes are excellent reference probes for comparison calibrations in a Hart dry-well.
A printout of sensor resistance is provided in 1°C increments for each probe. The 5614 and 5612 are calibrated from –196°C to 420°C. The 5613 is calibrated to 300°C.
Weve tested many of the probes on the market. Weve used them in our manufacturing facility and tested them in the lab, and this is an excellent secondary standards PRT. Other instruments on the market are priced much higher, have lower stability, or have lower quality.
Remember, these instruments are inexpensive and have excellent durability. Each probe is individually calibrated and includes a report of calibration from the manufacturer. Contact Alpha Controls for calibration in Harts NVLAP accredited lab.
|