Process Controls Home
 
 

X Cells - Stainless Steel Fixed-Point Cells

X Cells - Stainless Steel Fixed-Point Cells

X,Cell,Stainless,Steel,Fixed,Point,Cells

Features

  • Stainless steel casings help protect cells from mishandling
  • Accuracies virtually identical to traditional-size quartz cells
  • Can be maintained in fluid baths or dry-well calibrators

Specifications

Nominal Temperature Values

Water: 0.01°C
Gallium:
29.7646°C
Indium
: 156.5985°C

Expanded Uncertainty (k=2)

Water: ±1.0 mK (0.5 typical)
Gallium
: ±1.0 mK (0.5 typical)
Indium
: ±2.0 mK (1.0 typical)

Metal Sample Purity

Gallium: 99.99999%
Indium
: 99.9999%

Immersion Depth
(in pure sample)

Water: 3.4" (86.4 mm)
Gallium
: 3.0" (76.2 mm)
Indium
: 3.0" (76.2 mm)

Casing Material

Stainless Steel

Well I.D.

0.25" (6.35 mm)

Cell Size

5" H x 1" Dia.
(127 x 24 mm)

Have you wanted to use fixed-point cells but couldn’t justify the price? Have you been hesitant because of complex realization procedures or the fragile nature of traditional cells?

Hart’s new X Cells solve all these problems. They provide primary standards performance, are nearly unbreakable, and cost much less than traditional cell systems. The value of the water triple point X Cells is within 0.2 mK of traditional water triple point cells, and the gallium and indium cells include realized uncertainties less than 0.002°C.

Gallium and indium X Cells are constructed using a Teflon crucible containing a sample with purity of at least 99.9999%. The crucible is enclosed within a specially cleaned stainless steel envelope, which is evacuated and back-filled with high-purity argon.

Realizing these reference points couldn’t be easier. Special X Cell maintenance baskets allow the cells to be used in both standard baths and Micro-Baths, and a special sleeve is available for maintaining X Cells in Model 9103 dry-well calibrators.

Simply place the cell in the heat source, select the melt temperature, and within 30 minutes you’ll have an ITS-90 reference temperature all day long with stability of ±0.001°C. Once you finish using the cell, just set the freeze temperature and within minutes the cell is again ready for another realization. What could be easier?



Powerful Solutions for Instrumentation & Control Problems!

X Cells - Stainless Steel Fixed-Point Cells

Site Designed By Panther Marketing