Standard Resistor Baths

|
Resistor Baths Models 7008, 7015, and 7009
|
|
Three size options for any quantity of resistors
|
|
Stability to ±0.0007°C
|
|
Independent high- and low-temperature cutout circuit
|
|
Minimal long-term drift
|
|
Specifications
|
7008
|
7015
|
7009
|
| |
|
|
|
|
Range
|
5°C to 110°C
|
0°C to 110°C
|
0°C to 110°C
|
|
Stability at 25°C
|
±0.0007°C at 25°C (water)
±0.001°C at 25°C (oil)
|
|
Uniformity
|
±0.003°C at 25°C (water)
±0.004°C at 25°C (oil)
|
±0.003°C at 25°C (water)
±0.005°C at 25°C (oil)
|
±0.005°C at 25°C (water)
±0.007°C at 25°C (oil)
|
|
Temperature Setting
|
Digital display with push-button data entry
|
|
Set-Point Resolution
|
0.002°C; high-resolution mode, 0.00003°C
|
|
Display Resolution
|
0.01°C
|
|
Digital Setting Accuracy
|
±1°C
|
|
Digital Setting Repeatability
|
±0.01°C
|
|
Heaters
|
500 and 1000 Watts
|
|
Cooling Capacity
|
100 to 200 Watts
|
|
Access Opening
|
12.75" x 7.25"
(324 x 184 mm)
|
27.5" x 11"
(699 x 279 mm)
|
27.5" x 22"
(699 x 559 mm)
|
|
Bath Chamber Dimensions (unobstructed space)
|
14.25" W x 10.75" H x 13" D
(1991 cubic inches)
|
27.5" W x 11" H x 13" D
(3933 cubic inches)
|
27.5" W x 22" H x 13" D
(7865 cubic inches)
|
|
Depth
|
13" (331 mm)
|
|
Wetted Parts
|
304 stainless steel
|
|
Safety Cutout
|
Factory-set high and low temperatures
|
|
Power
|
115 VAC (±10%), 60 Hz, 15 A or 230 VAC, 50 Hz, 8 A, specify, 1700 W
|
230 VAC (±10%), 50 or 60 Hz, 12 A, 2600 W (specify frequency)
|
|
Volume
|
11.2 gallons
(42 liters)
|
25 gallons
(95 liters)
|
44 gallons
(167 liters)
|
|
Weight
|
135 lb. (61 kg)
|
310 lb. (141 kg)
|
330 lb. (150 kg)
|
|
Size
|
24" H x 30.5" W x 19" D
(610 x 775 x 483 mm)
|
48" H x 44" W x 22" D
(1219 x 1118 x 559 mm)
|
43" H x 44.5" W x 34" D
(1092 x 1130 x 864 mm)
|
|
Automation Package
|
Interface-it software and RS-232 computer interface are available for setting the bath temperature via an external computer. For IEEE-488, add the 2001-IEEE to the automation package.
|
Regardless of the size and number of standard resistors you have to maintain, Hart has a bath that will do the job for you. Choose one of the three models described here or call us for information on other sizes.
Like all Hart baths, these resistor baths have unbeatable stability and uniformity. No other baths limit long-term and short-term driftas well as gradientsbetter than these baths. Harts proprietary controller senses temperature changes as small as 0.00001°C. This controller is the industrys best-selling temperature calibration controller for bath retrofits because it improves the stability of almost every other poorly performing bath. So why not buy the best to begin with?
Each bath can be delivered with any size resistor rack you want (a standard model is included with each bath), and the Model 7015 has several other special features that make your work easier. A new independent high- and low-temperature cutout provides maximum protection for your standard resistors and their calibration history.
Model 7015
The Model 7015 has a 95-liter tank and a temperature range of 0°C to 110°C. Its stable to ±0.0007°C.
It has a one-piece stainless steel lid designed to drain spills and splashes back into the bath as you remove resistors. It has a large access opening to make handling large resistors, like the Thomas Design Standard resistors, easier. The tank has an electrically isolated resistor shelf.
This is truly a quality resistor bath, and its backed by Harts industry-leading service.
Model 7008
This 44-liter bath has a range of 5°C to 110°C and a stability of ±0.0007°C. Its available with an RS-232 port or IEEE. There are cart options and oil for any application.
|