Process Controls Home
 
 

Rotronic MPH Series Specifications

Technical data

MP100H (voltage outputs)

MP400H (current outputs)

Configurations
  • Separate temperature probe with active linear output signal (standard)
  • Separate temperature probe with direct 2-, 3- or 4-wire RTD connection (optional)
  • No separate temperature probe - temperature output provided by RTD from HygroClip (optional)
Humidity Temperature Plug-in Probe HygroClip S3 (included with MPH)
Humidity sensor: ROTRONIC Hygromer® C94 (capacitive)
Temperature sensor: RTD Pt100 Ohm, 1/3 DIN
Sensor protection: Wire mesh filter (standard)
                            Foam filter (optional)
Separate Temperature Probe Stainless steel rod 118 x 3 mm (4.64 x 0.12")
Pt100 1/3 DIN RTD (standard)
Pt100 1/5 DIN RTD (optional)
Pt100 1/10 DIN RTD (optional)
Operating/Measuring range 0...100%RH
-40...60°C / -40...140°F
Accuracy at 23°C / 73°F <±1.5%RH and ±0.2°C (valid for active outputs)
Repeatability better than 0.5%RH and 0.1°C
Measuring ranges
(active outputs)
0...1V = 0...100%RH 4...20mA = 0...100%RH
0...1V = -40...60°C (standard)
0...1V = -30...70°C (optional)
-0.4...0.6V = -40...60°C (optional)
4...20mA = -40...60°C (standard)
4...20mA = -30...70°C (optional)
Active Outputs Load > 1000 Ohm < 500 Ohm
Supply Voltage 3.6...35VDC, 10 mA 5VDC + 0.02 x load, 45 mA max.
Cable Length Compensation effective for up to 100m/300ft N/A (current outputs)
Connecting Cable standard: 3m (9ft) PUR cable with tinned ends
Weight 210 g / 0.47 lb


Powerful Solutions for Instrumentation & Control Problems!
Rotronic MPH Series Specifications

Site Designed By Panther Marketing