HW3 is a software package designed for the HygroClip product line. HW3 features data acquisition, recording, graphing and alarm functions, calculation of various humidity parameters (both on-line and off-line) as well as instrument configuration and probe calibration. HW3 also offers a comprehensive solution for the networking of up to 32 HygroFlex M3 transmitters.


Rotronic,HW3,Software
Network Display Form

• On-line presentation of the data in tabular form
• Up to 32 transmitters
• Up to 64 humidity and temperature probes
• Up to 192 measured and calculated values
• Live graph of up to 8 measured or calculated values
  with or without data logging
Data-logging

• Log interval adjustable from a few seconds to several hours
• Up to 32 instruments can be simultaneously recorded
• Individual log file per instrument or all instruments on the network
  in a single log file (requires Microsoft's Excel)
• Size of log file can be limited based on the number of samples or a new file
  can be automatically started each hour, day or month
• Display or print-out of the data in tabular format
• File data can be transferred to HW3 graphic module or to Microsoft's Excel
Alarm function

• Monitors relative humidity and / or temperature
• Can be set for individually for each instrument and probe on the network
• Acoustic and or visual signal directly on PC
• Remote alarm by email (requires Microsoft's Outlook)
Graphs

• On-line (live graph) or off-line (based on a log file)
• Can display calculated values: dew-point, mixing, ratio etc.
• Up to 16 traces simultaneously
Instrument Display Form

• On-line presentation of the data for a single transmitter
  (number of probes depends on the model and its options)
• Selection of the calculated parameter
• Access to data logging, graph and alarm functions
• Access to instrument configuration and probe calibration


Probe Calibration (HygroClip probes)

• Humidity: up to 4 points, Temperature: 2 points
• 1-point adjustment function (offset)
• Automatic temperature correction of the ROTRONIC humidity standards
• Certificate can be generated


System Requirements


PC




Pentium 233 MHz or better
Windows 95, 98, Me or NT, 2000
Memory 32 MB or better
Disk Space 20 MB
Monitor VGA or Super VGA
Resolution 1024 x 768 recommended (800 x 600 minimum)
High Color 16 Bit or better (256 color minimum)
Connection RS232 (COM 1 - 4)
Software HW3 (includes Browser, Adobe Acrobat Reader)


Powerful Solutions for Instrumentation & Control Problems!
Rotronic HW3 Software

Site Designed By Panther Marketing