HygroLab Multi-Channel Humidity/Temperature
And Water Activity Indicator
|
 |
Models | Probes | Software | Calibration
Multiple Probe Inputs
Menu
Display
|
 |
Enter
ON/OFF
Up/Down
|
Developed for busy laboratory environments where multiple measurement stations or probes co-exist, the HygroLab 2 features networking, allowing up to 32 instruments to be simultaneously controlled and monitored from a single PC.
Based on the very latest digital technology, HygroLab offers the user outstanding performance and flexibility, with up to four simultaneous probe inputs and an impressive number of functions.
Probes can be interchanged without loss of accuracy. The HygroLab features a full range of psychrometric calculations, software based measurement analysis and multi-point calibration.
An extensive range of probes and water activity measuring stations make the HygroLab suitable for almost any laboratory or production application.
HygroLab is the new series of bench top indicators for the measurement of humidity and water activity (aw).
Models
 |
|
HygroLab 1
Key Features
Two digital or analogue probes
One point digital probe calibration |
 |
|
HygroLab 2
Key Features
Four digital or analogue probes
Single or multi-point digital probe calibration
Full psychrometric calculation functions
Networkable: RS232/RS485 output |
Probes
|
HygroClip S
for measurement in air, max. 85ºC (185ºF), wire mesh filter
100mm (3.9") x 15mm (0.6") |
|
HygroClip SP05
for measurement in air ducts, max. 85ºC (185ºF)
tube: 200mm (7.9") x 5mm (0.2")
cable: 2m (6.5ft) |
|
HygroClip SC05
for measurement in tight spaces, max. 100ºC (212ºF)
diameter: 5mm (0.2"), cable: 2m (6.5ft) |
|
HygroClip HK25 / HK40
air probe for measurement at high temperature
HK25: max. 100ºC (212ºF) / HK40: max. 200ºC (392ºF)
wire mesh filter
tube:HK25: 250 (9.8") / HK40: 400mm (15.7") x 15mm (0.6")
cable: 2m (6.5ft) |
|
HygroClip HP28 insertion probe
for measurement of materials in bulk, max. 85ºC (185ºF)
sintered steel filter
tube: 280 (11") x 10mm (0.4")
cable: 2m (6.5ft) |
|
HygroClip HS28 sword probe
for measurement of paper stacks/rolls, max. 85ºC (185ºF)
blade: 280 (11") x 18mm (0.7") x 4mm (0.16")
cable: 2m (6.5ft) |
To connect the above probes to the HygroLab, use cable MOK-01-B5
|
 |
|
AwVC-DIO water activity station with ventilation for rapid equilibrium |
 |
|
WP-40TH sample holder with integrated water jacket for temperature control and PS-40/PS-14 sample cups |
 |
|
WP-40/WP-14 sample holder shown with PS-14/PS-40 disposable sample cups |
Software
| HW3 is a software package designed for the HygroClip product line. HW3 features data acquisition, recording, graphing and alarm functions, calculation of various humidity parameters (both on-line and off-line) as well as instrument configuration and probe calibration. HW3 also offers a comprehensive solution for the networking of up to 32 HygroLab 2 indicators and transmitters. |
 |
Network Display Form
On-line presentation of the data in tabular form
Up to 32 instruments
Up to 128 humidity and temperature probes
Up to 384 measured and calculated values
Live graph of up to 8 measured or calculated values
with or without data logging
|
 |
Data-logging
Log interval adjustable from a few seconds to several hours
Up to 32 instruments can be simultaneously recorded
Individual log file per instrument or all instruments on the network
in a single log file (requires Microsoft's Excel)
Size of log file can be limited based on the number of samples or a new file
can be automatically started each hour, day or month
Display or print-out of the data in tabular format
File data can be transferred to HW3 graphic module or to Microsoft's Excel
|
 |
Alarm function
Monitors relative humidity and / or temperature
• Can be set for individually for each instrument and probe on the network
Acoustic and or visual signal directly on PC
Remote alarm by email (requires Microsoft's Outlook)
|
 |
Graphs
On-line (live graph) or off-line (based on a log file)
Can display calculated values: dew-point, mixing, ratio etc.
Up to 16 traces simultaneously
|
 |
Instrument Display Form
On-line presentation of the data for a single instrument
(up to 4 probes per instrument)
Selection of the calculated parameter
Access to data logging, graph and alarm functions
Access to instrument configuration and probe calibration
|
 |
Probe Calibration (HygroClip probes)
Humidity: up to 4 points, Temperature: 2 points
1-point adjustment function (offset)
Automatic temperature correction of the ROTRONIC humidity standards
Certificate can be generated
|
System Requirements
PC |
Pentium 233 MHz or better |
| Windows |
95, 98, Me or NT, 2000 |
| Memory |
32 MB or better |
| Disk Space |
20 MB |
| Monitor |
VGA or Super VGA |
| Resolution |
1024 x 768 recommended
(800 x 600 minimum)
High Color 16 Bit or better
(256 color minimum) |
| Connection |
RS232 (COM 1 - 4) |
| Software |
HW3 (includes Browser, Adobe Acrobat Reader) |
|
The digital HygroClip probes can be calibrated and adjusted at one or more points with the HygroPalm 2 or HygroPalm 3 indicator:
For a quick check of the probe and eventual 1-point adjustment of humidity and temperature, you can plug a freshly calibrated probe into any free probe input and adjust against this probe.
For scheduled maintenance, you can adjust temperature at up to 2 points and humidity at up to 4 point against physical references.
Note: the HygroPalm 1 is limited to a 1-point adjustment of the probe. The HygroPalm 2 and 3 permit full calibration of the probe.
You may also make use of the full interchangeability of the HygroClip digital probes and rotate the probes with a calibration laboratory. The calibration laboratory can do any of the following:
|
Powerful Solutions for Instrumentation & Control Problems!
HygroLab Multi-Channel Humidity/Temperature And Water Activity Indicator
|
|