Process Controls Home
 
 

Capacitance Vacuum Gauge Setra Model 720 CVG

This low cost, compact vacuum gauge is designed to be used with pressure media compatible with Inconel.® The accuracy of this transducer is assured with a wide compensated operating temperature of 0 to +50°C and exceptional temperature coefficients of ± 0.005% FS/°C at Zero and ± 0.027% of Rdg/°C at span.

Capacitance,Vacuum,Gauge,Setra,Model,720,CVG

Applications

  • Semiconductor
  • Absorption Chillers
  • Lasers
  • Autoclaves
  • Vacuum Packaging
  • Freeze Drying
  • Vacuum Distillation
Model 720 Specifications
Electrical Excitation/Output

14 to 30 VDC for 0 to 10 VDC
9 to 30 VDC for 0 to 5 VDC

Electrical Termination

9 Pin D-Sub

Accuracy ±0.5% of Reading or ±0.25% of Reading (optional)
Type of Pressure Absolute
Pressure Ranges 10, 20, 100, 1000 Torr
10, 20, 100, 1000 Millibar/hPa
1, 2, 10, 100, kPa
Thermal Effects
Compensated Range °F (°C):
0 to +50 (+32 to +122)
Zero Shift: ±0.005% FS/°C 
Span Shift: ±0.027% Rdg/°C 
Media Gases or liquids compatible with Inconel®  Inconel® wetted material is for 0.5" tube option only. Other fitting options will add stainless steel.

 

* Specifications subject to change without notice.
If you don't see what you need here, please contact us.


Powerful Solutions for Instrumentation & Control Problems!

Capacitance Vacuum Gauge Setra Model 720 CVG

Site Designed By Panther Marketing