| Measuring Range |
Type K : -190ºC to 1333ºC (-310ºF to 2431ºF)
Type J : -190ºC to 760ºC (-310ºF to 1400ºF) |
| Accuracy |
(-190ºC to -0ºC ) ( +/- 0.5% rdg + 0.7ºC)
(-310ºF to -0ºF ) ( +/- 0.5% rdg + 1.4ºF)
( 0ºC to 1000ºC) ( +/- 0.1% rdg + 0.5ºC)
( 0ºF to 1000ºF) ( +/- 0.1% rdg + 1.0ºF)
( 1000ºC to 1333ºC) ( +/- 0.2% rdg + 1ºC)
( 1000ºF to 2431ºF) ( +/- 0.2% rdg + 2ºF) |
| Resolution |
-190ºC to 999ºC (0.1ºC)
1000ºC to 1333ºC ( 1ºC)
-310ºF to 999ºF (0.1ºF)
1000ºF to 2431ºF ( 1ºF) |
| Input Protection |
60VDC/24Vrms |
| Power Source |
One 9V battery. 006p or NEDA1604 or IEC6F22 |
| Battery Life |
Approx. 50hr |
| Operating / Storage Condition |
0ºC to 40ºC (32ºF to 104ºF)below 80¢MRH.
-10ºC to 60ºC (14ºF to 140ºF) below 70¢MRH. |
| Dimensions |
145(L)¡Ñ68(W)¡Ñ35(H)mm |
| Weight |
Approx. 235g with battery |
| Accessories |
Instruction manual, Battery, Carry case, SºFtware,
RS-232 cable, 9 Pin to 25 Pin gender changer. |