Models 8330 and 8340
TSI's line of VELOCICHECK Air Velocity Meters provides a reliable, accurate and easy way to make air velocity measurements. The VELOCICHECKs are small, pocket-sized meters that require no training; just pick up the meter and start taking readings. These meters
provide a cost-effective way to quickly take the measurements you need.
Features
Specifications
Downloads
(Brochures, Manuals, etc. Use your browser's back button to return)
Model 8330 Features
- Extended velocity range of 0 to 20.00 m/s
- Temperature range of -18 to 93°C
- Easy-to-read digital display
- Slow and fast time constant modes available to smooth the display when measuring fluctuating flows
- 94.0 cm telescoping probe provides exceptional convenience in difficult situations
- Includes a battery check that monitors battery level to ensure accurate readings
- NIST*-traceable calibration certificate
Model 8340 Features
- Listed by Underwriters Laboratories as intrinsically safe for use in hazardous locations for
Divisions 1 and 2 in Class I, Groups C and D; Class II, Groups E, F and G; and Class III.
- Velocity range of 0 to 10.00 m/s
- Easy-to-read digital display
- Slow and fast time constant modes available to smooth the display when measuring fluctuating flows
- Includes a battery check that monitors battery level to ensure accurate readings
- NIST*-traceable calibration certificate

Specifications - Models 8330 and 8340 VELOCICHECK
Velocity (Model 8340):
Range: 0 to 10.00 m/s (0 to 2,000 ft/min)
Accuracy1: 5.0% of reading or ±0.025 m/s (±5 ft/min) whichever is greater
Velocity (Model 8330):
Range: 0 to 20.00 m/s (0 to 4,000 ft/min)
Accuracy2: 5.0% of reading or ± 0.025 m/s (±5 ft/min) whichever is greater
Temperature (Model 8330):
Range: -18 to 93°C (0 to 200°F)
Resolution: 1°C (1°F)
Accuracy: ±1°C (±1°F)
External Meter Dimensions (all models):
Size Measurements: 70 mm x 33 mm x 132 mm (2.75 in. x 1.3 in. x 5.2 in.)
Meter Probe Type (Model 8330):
Description: Detached telescoping
Material: Nickel-plated brass and stainless steel
Meter Probe Type (Model 8340):
Description: Retractable
Material: Stainless steel
Meter Probe Dimensions (Model 8330):
Probe Length: 94.0 cm (37 in.) telescopic
Probe Diameter of Tip: 6.0 mm (0.236 in.)
Probe Diameter of Base: 10.03 mm (0.395 in.)
Meter Probe Dimensions (Model 8340):
Probe Length: 79 mm (3.1 in.) length when extended
Probe Diameter: 6.4 mm (0.25 in.)
Operating Temperature Range (Model 8330):
Instrument: 0 to 60°C (32 to 149°F)
Probe: -18 to 93°C (0 to 200°F)
Operating Temperature Range (Model 8340):
Instrument: 0 to 50°C (32 to 122°F)
Probe: 0 to 50°C (32 to 122°F)
Battery Life:
Model 8330: Minimum 10 hours at 0.5 m/s (100 ft/min)
using alkaline batteries
Model 8340: Minimum 20 hours at 0.5 m/s (100 ft/min) using alkaline batteries
Display (all models):
Type: Digital
Description: 4 1/2 digits, 10 mm (0.4 in.) high
Time Constant (all models):
Fast mode: 3-second averaging
Slow mode: 12-second averaging
Power Requirements:
Model 8330: Four AA-size batteries, nickel cadmium or alkaline (provided)
Model 8340: Four AA-size batteries, alkaline only (provided)
Safety Rating (Model 8340): UL listed intrinsically safe for use in hazardous locations for Divisions 1 and 2 in Class I, Groups C and D; Class II, Groups E, F and G; and Class III when used with alkaline batteries
Standard Equipment (all): Free certificate of NIST* traceability, operation and service manual, four size AA alkaline batteries and hard-shell carrying case
|
Velocity
Range
0 to 10.00 m/s |
Velocity
Range
0 to 20.00 m/s |
Temp. |
Digital
Display |
Time
Constant |
NIST*
Calibration
Certificate |
Intrinsically
Safe |
| 8330 |
|
 |
 |
 |
 |
 |
|
| 8340 |
 |
|
|
 |
 |
 |
 |
*U.S. National Institute of Standards and Technology
1The accuracy statement of 5.0% of reading or ±.025 m/s (±5 ft) applies to 0.13 through 10.00 m/s (25 ft/min through 2000 ft/min).
2The accuracy statement of 5.0% of reading or ±.025 m/s (±5 ft) applies to 0.13 through 20.00 m/s (25 ft/min through 4000 ft/min).
Specifications are subject to change without notice.
Specifications in parentheses ( ) indicates metric equivalents.
|