Universal Measurement System for Gas-Exchange Metabolism based on the manometric measurement of pressure changes (respirometer) in a closed system
Applications
aerobic
- determination of soil respiration (DIN 19 737)
- AT4-determination oxygen consumption (cell respiration; composting, BOD-determination)
- determination of biodegradability OECD 301 F
- oxygen absorption by solid matter, etc.
anaerobic
- biogas build-up (fermentation)
- gas discharge (waste dumps, sewage sludge, etc.)
aerobic and anaerobic
- quantifying the CO2 content of accumulated biogas
- pressure differences, changes in pressure patterns, etc.
|
Measurement ranges
|
Pressure p operating mode: 500 ... 1350 hPa (=mbar) Standard & Routine BOD operating modes: 0 ... 40 / 80 / 200 / 400 / 800 / 2000/ 4000 mg/l BOD Special BOD operating mode: 0 ... 400,000 mg/l BOD;
|
|
Run times & data sets Measurementperiod
|
Run time of measurement 0.5 hours 1.5 hours 2.5 hours 3.5 hours 4.5 hours 5.5 hours
|
Data records per Sensor-IR 180 270 300 315 324 330
|
|
all whole hours
|
(1, 2, 3 to 99)
|
360
|
|
all whole days
|
(1, 2, 3 to 23)
|
360
|
|
Display
|
LCD graphics display 64 x 128 pixels
|
|
Memory capacity
|
Sensomat-Scientific: Data of 120 Sensor-IR
|
|
Power supply
|
Supply batteries: 3 pcs, alkaline (alkaline manganese), Size: Mignon, AA, AM3, LR6
Data backup battery: 1 x lithium battery, CR2430, Cat.-No. 427 608
|
|
Battery run time
|
Supply batteries:>1000 switchings with normal use; (the instrument has an automatic shut off)
Data backup battery: Typically 4 years, Cat.-No. 418 608
|
|
Safety class
|
3, equipment safety: IEC 1010, EN 61010 Teil 1
|
|
Protection type
|
IP 41 DIN 40050
|
|
EMC
|
Interference emission: EN 50081-1, FCC Class A
Interference immunity: EN 50082-1
|
|
Climate class
|
2, VDI/VDE 3540
|
|
Ambient temperature
|
Storage: -25¡C ... +65¡C Operation: +5¡C ... +40¡C
|
|
Dimensions
|
45 x 100 x 200 mm (H x W x D)
|
|
Weight
|
approx. 390 g
|
|
Test certificates
|
CE, T†V/GS, UL, CUL (UL/CSA)
|
|
Year-2000-comp.
|
yes
|
|