- measurements from up to 100 different measurements points can be accessed simply by push-button with infra-red remote operation
- duration of measurement can be selected between 30 minutes ... 99 days: e.g., BSB 5 , BSB 7 , BSBn or short-term BOD measurements
- a really user-friendly means of BOD measurement
- graphic presentation of BOD curves with mean value display
- software-supported evaluation of series of measurements with statistical functions
- electronic storage ensures extra security for measurements and other data
- complies with GLP-/ISO 9000
The BSB/BOD-Sensomat-System comprises three interlocking, non-contact modules:
- the pressureSENSOR-IR acts simultaneously as the measurement device and the storage unit for the measured results. Together with the sample vessel, it forms a mobile, self-contained measuring unit.
- theBSB/BOD-Sensomat is the transportable §core" of the system: it logs all measured data at the push of a button and ensures simple, comprehensive evalutation and documentation of results.
- theInductive Stirring System ensures the gas-exchange between the water sample and the air phase by periodic monitoring of the centering of the magnetic agitation rod.
|
Measuring principle:
|
Respirometric; mercury-free; electronic pressure sensor
|
|
Measuring ranges [mg/l O2]:
|
0 - 40, 0 - 80, 0 - 200, 0 - 400, 0 - 800, 0 - 2000, 0 4000 mg/l
|
|
Applications:
|
BOD5, BOD7, OECD 301 F ...
|
|
Result display:
|
BOD [mg/l]; 4 digit; 7-segment LED
|
|
Measuring parameter display:
|
BOD, volume, duration, measuring time
|
|
Measuring duration:
|
Selectable, between 1 and 28 days
|
|
Automatic result memory:
|
Up to 28 results, depending on measuring duration
|
|
Storage interval:
|
Hourly to daily, depending on measuring duration
|
|
Autostart:
|
on/off switching facility, measure-ment begins after pressure drops, or maximum waiting period
|
|
Current result:
|
Can be called up at any time
|
|
Power supply:
|
3 alkaline manganese C (Baby) batteries
|
| Battery life: |
1 year based on normal use as BOD5 measuring unit (max. one reading daily); early warning at end of battery life |
|
Clock:
|
Real-time clock
|
|
Protection class (sensor head):
|
IP54
|
|
Housing:
|
ABS
|
|
Approval certificate:
|
CE
|
Measuring Systems
|
Cat.-No.
|
Item
|
|
418336
|
BSB/BOD-Sensomat-System AL 336 for 6 measuring places, complete with BSB/BOD-Sensomat, Sensor-IR, Inductive Stirring System AL 311 and standard accessories, 230V/50-60 Hz
|
|
418347
|
BSB/BOD-Sensomat-System AL 336 for 6 measuring places, complete with BSB/BOD-Sensomat, Sensor-IR, Inductive Stirring System AL 311 and standard accessories, 115V/50-60 Hz
|
|
418342
|
BSB/BOD-Sensomat-System AL 342 for 12 measuring places, complete with BSB/BOD-Sensomat, Sensor-IR, Inductive Stirring System AL 322 and standard accessories, 230V/50-60 Hz
|
|
418348
|
BSB/BOD-Sensomat-System AL 342 for 12 measuring places, complete with BSB/BOD-Sensomat, Sensor-IR, Inductive Stirring System AL 322 and standard accessories, 115V/50-60
|
|