- pH / redox measurement and temperature measurement
(Pt1000 or NTC 30kOhm)
- Automatic temperature compensation (ATC)
- Auto Hold function
- Internal memory for 20 data sets
- Automatic buffer recognition
- 1, 2 and 3 point calibration
- Low bat and battery change indicator
- Sensor evaluation in the display
The new microprocessor-controlled SensoDirect hand-held meter range from AQUALYTIC® meets the day-to-day demands on sturdy and reliable systems for the measurement of pH or redox, conductivity/TDS and dissolved oxygen.
The water-tight housing in compliance with IP67 and equipped as standard with protective armouring ensures reliable operation even in extreme ambient conditions. The support can be flipped up to hang the meter on pipes or branches.
The direct, convenient and easy-to-understand user prompt with the required configuration options for all three systems facilitates the use of the meters both outdoors and in the lab.
The automatic Hold function freezes stable measuring data in the display and indicates to the user the presence of stable and reproducible results.
The internal memory permits storage of 20 data sets to facilitate subsequent evaluation. The integral auto switch-off feature that can be set from 1 to 120 minutes, increases the operating life of the units.
The power consumption of all three unit versions has been reduced to a minimum. As a result, the 4 x 1.5 V integrated batteries are good for an operating life of up to 10,000 hours, depending on the unit version.
AQUALYTIC® Electrodes/Measuring Probes
A wide variety of high-quality electrodes and measuring probes for the SensoDirect meters make the units suitable for almost any application in industry and research.
Single-rod measuring chains made of glass and plastic, with gel or liquid electrolyte, as well as various temperature sensors (NTC 30kOhm, Pt1000) are considered standard for modern pH/ORP Process Controllers.
Redox electrodes made of plastic with gel electrolyte and platinum tip are hardwearing and easy to use. The amperometric, diaphragm-covered oxygen sensor with integrated NTC allows instant measurement without the need for time-consuming polarisation.
The sturdy conductivity measuring cell with two graphite electrodes is ideal for use in the field of service and waste water analysis.
|
Indicating ranges:
|
|
Temperature:
|
-10.0 ... + 110,0 C
|
|
pH:
|
0,00 ... 14,00 pH
|
|
Redox (ORP)
|
-1999 ... + 2000 mV -1792 ... +2206 Rel mV (DIN 38 404)
|
|
|
Accuracy:
(at nominal temperature unit ± 1 digit)
|
|
Temperatur:
|
±0.2 °C (-0..50 °C), otherwise ± 0.4 °C
|
|
pH:
|
± 0.01 pH
|
|
Redox (ORP)
|
± 0.1 % FS (mV or Rel mV)
|
|
|
Temperature compensation:
|
With temperature probes, automatic temperature compensation (ATC) is effected in measuring modes pH or Rel mV. Without measuring probe, manual temperature input is necessary.
|
|
Sensor connections:
|
Temperature: two 4 mm banana sockets Temp und Ref Temperature probes PT1000 and NTC30k pH, redox: DIN19 262 socket, separate reference electrode is connected to Ref
|
|
Input resistance:
|
>1012 Ohm (typ.)
|
|
Input current:
|
<1 pA
|
|
Nominal temperature:
|
25 °C
|
|
Operating temperature:
|
0 to +50 °C
|
|
Relative humidity:
|
0 to +95 % rel. hum. (no condensation)
|
|
Storage temperature:
|
-20 to +70 °C
|
|
Memory:
|
for 20 results
|
|
Power supply:
|
4 x 1.5 V battery, type AA operating time up to >10,000h (depending on measuring mode and battery type)
|
|
Power consumption:
|
max. 160 µA
|
|
Automatic Off function:
|
If no key is pressed or if there is no data transfer, the unit switches off after run-down of the switch-off delay. Settable from 1 - 120 minutes or deactivated.
|
|
Housing:
|
Water-tight in acc. with IP67, Integrated protective armouring, Liquid crystal display (LCD), Film keypad
|
|
pH calibration:
|
1-point, 2-point or 3-point calibration Standard buffer, DIN buffer or user-selectable buffer
|
|
Auto buffer recognition:
|