The
advanced "isolated" circuitry of the 720 Series
II pH/ORP Monitor/controllers guarantees accurate and reliable
measurements completely eliminating ground-loop and noise
issues.
The unique sensor preamp allows for longer distances
between the sensor and the Monitor/controller without the
loss of accuracy or reliability.
All Myron L pH Monitor/controllers feature a
highly refined and precise Temperature Compensation circuit.
This feature perfectly matches the NERNST equation correcting
the displayed reading to 25ºC. The TC may be disabled
to conform with USP 24.
Built-in electronic calibration allows for fast
quality checks without buffers.
(Note: for maximum system accuracy buffers are recommended).
A push-to-test button makes set point checks
easy, while the internal alarm set discourages unauthorized
adjustments.
Standard on all controller models is a heavy-duty
10 amp output relay, operating on either increasing or decreasing
readings. Bright green/red LED's indicate above/below
set point readings.
All 720 Series II models feature a IPA64/NEMA
3 splash & corrosion-proof rated enclosure. Compact
in size, 6 x 4.8 inches (152 x 122mm), the 720 Series II models
are suitable for panel, bench or surface mounting.
Accessories include a Remote Alarm - RA™.
Custom and OEM Models available.
pH/ORP
Sensors
These highly accurate sensors utilize a preamplifier to ensure
stability over longer distances, and to eliminate potential
noise problems thereby ensuring accurate repeatable readings.
In addition, these reliable sensors feature
an integral temperature sensor to ensure accurate and rapid
automatic temperature compensation.
Several choices of sensors are available, in
1/2 MNPT or 3/4 MNPT, depending on the demands
of the application. The basic pH sensor utilizes the standard
round glass electrode with a single reference junction. The
basic ORP sensor utilizes a Platinum electrode with a single
reference junction. See 720II Sensor Selection Guide for additional
information.
NOTE: Heavy Duty F model sensors
may be installed in ANY direction including inverted.
Applications
- Process Control
- Reverse Osmosis
- Seawater Desalination
- Waste Treatment
- Food Processing
- Plating
- Power Plants
- Laboratories
- Printing
- Paper & Pulp
- Boilers & Cooling Towers
- Potable Water
- Waste Water Treatment
- Swimming Pools & Spas
- Bleach Manufacturing
- Reduction of Chromate Waste
- Environmental
Features
- pH 0-14 or ORP ± 2000 mV
- Choice of 3½ digit display or analog meter
- Advanced "Isolated" circuitry
- Automatic temperature compensation to 25°C (NERNST
equation)
- Isolated 0-10 VDC Recorder Output standard
- IP65/NEMA 4X water resistant and corrosion-proof enclosure
- Built-in electronic calibration
- Single set point/alarm standard
- Adjustable set point hysteresis
- Solid State 24 VDC Output
- 10 amp relay operates on choice of increasing or decreasing
reading
- Push-to-test set point check
- Internal set point discourages unauthorized adjustment
- Selectable 115/230 VAC
- Unique, easily installed sensors
- CE Certified
- Unique, easily installed sensors
- Heavy duty F model sensors may be installed
in ANY direction including inverted
Sensor
Specifications
- Temperature Compensation: Integral temperature sensor
provides automatic TC to 25°C between 0 - 100°C/
32 - 212°F.
- Pressure/Temperature Limits: P72S: Intermittent use
up to 50 psi/690 kPa 100°C/212°F. High Temperature
models available, see pH/ORP Sensor Selection Guide.
- Cable: 5 wire, 24 ga. shielded, 10 ft./3 meters standard.
25 and 100 ft./8 and 30 meter lengths available.
- Material: 1/2" MNPT: CPVC body. 3/4" MNPT:
Ryton®* body.
- Dimensions: P72S typical: overall length 165 mm/6.5
in., body 127 mm/5 in., electrode length 19 mm/.75 in.