Resistivity,Monitors,Controllers

Resistivity Monitors Controllers


Resistivity Monitors Controllers

The unique circuitry of our 750 Series II Resistivity Monitor/Controllers guarantee accurate and reliable measurements.  Drift free performance is assured by "field proven" electronics, including automatic DC offset compensation and highly accurate drive voltage.

Since temperature compensation is at the "heart" of accurate water measurement, all Myron L Monitor/controllers feature a highly refined and precise TC circuit.  This feature perfectly matches the water temperature coefficient as it changes.  All models corrected to 25°C.  Uncompensated models available to conform with USP 24.

A push-to-test button makes setpoint checks easy, while the internal alarm set discourages unauthorized adjustments.

Standard on all controller models is a heavy-duty 10 amp output relay, operating on either increasing or decreasing readings.  Bright green/red LEDs indicate above/below setpoint readings.

All 750 Series II models feature a IP64/NEMA 3X splash and corrosion proof rated enclosure while being a compact size at 6x4.8 inches (15.2x122mm).  They are suitable for panel, bench or surface mounting.

Additional options include a 3 Sensor input, a 3 Range switch (both installed separately from the monitor/controller), and a RA™ - Remote Alarm.  Custom and OEM Models available.  Contact us with your requirements.

Applications

Deionization (DI) and distillation ultrapure water treatment systems

Electronics
Pharmaceuticals
Laboratories
Food Processing
Plating
Power Plants
Boiler Condensate
Process Control


Features

Seven Ranges:  0-200 Kilohms to 0-20 Megohm-cm
Choice of analog meter or 3.5 digit LCD
Automatic temperature compensation to 25°C
IP64/NEMA 3X splash & corrosion proof enclosure
Single setpoint/alarm standard
Adjustable setpoint hysterisis 
Solid state output
10 amp relay operates on choice of increasing or decreasing reading
LED's show "above/below" alarm/control status
Push-to-test set point check
Internal set point discourages unauthorized adjustment
Unique, easily installed sensor
Selectable 115/230VAC

Resistivity Monitor/controller Specifications

Ranges Resistivity: 7 ranges from 0-200Kilohm to 0-20Megohm
Readout
Models 751II & 752II - 2½" (63mm) analog meter
Models 753II & 754II - 1/2" (13mm) 3½ digit LCD; opt. 4½ digit backlit LCD
Accuracy
Models 751II & 752II - ±2% of full scale
Models 753II & 754II - ±1% of full scale
Sensitivity 0.05% of full scale
Stability 0.05% of full scale
Repeatability 0.1% of full scale
Calibration Check Built in
Voltage Output 0-10 VDC @ 5 mA max. (linear); standard all models
Sensor Inputs 1 (3 input option available)
Control Function
Single alarm/controller - fully adjustable 0-100% of full scale
Second alarm/controller - with above specs, opt. on 752II & 753II
Hysterisis - Adjustable from 0.3-3% of full scale
Indicators - Above (green) and below (red) setpoint LEDs (reversible)
Contact Rating - SPDT 10 amp @ 250 VAC, 30 VDC.  Relay operates with increasing or decreasing reading (user selectable)
Solid State Output Powered; 24VDC unregulated 30mA max.
Power 115/230 VAC, 50/60 Hz, ±15% (user selectable)
Maximum Power Consumption 100mA @115VAC
Dimensions 6" (152mm) H x 4.8" (122mm) W x 3.8" (96mm) D
Construction Fully gasketed heavy-duty ABS case; splash & corrosion proof IP64/NEMA 3X rated
Ambient Temperature Range 32°F (0°C) to 140°F (60°C)
Weight 2 lbs (0.9 kg)