| Measures |
Dissolved oxygen, °C |
| % saturationRange
|
0.00 to 200% |
| % saturationResolution
|
1% |
| % saturationAccuracy
|
±1.5% full-scale |
| mg/l (ppm)Range
|
0.0 to 20.00 mg/l (ppm) |
| mg/l (ppm)Resolution
|
0.01 mg/l (ppm) |
| mg/l (ppm)Accuracy
|
±1.5% full-scale |
| TemperatureRange
|
0 to 50°C |
| TemperatureResolution
|
0.1°C |
| TemperatureAccuracy
|
±0.5°C |
| CalibrationDissolved oxygen
|
one or two points (100% and 0% saturation) |
| CalibrationTemperature
|
offset up to ±5°C |
| Control type
|
on/off or proportional (selectable) |
| Number of inputs
|
one |
| Number of set points
|
two (high and low) |
| Separate alarm relay
|
yes |
| OutputControl
|
2 SPDT relays, 6 A at 110 VAC, 250 VAC maximum |
| OutputAlarm
|
1 SPDT relay, 6 A at 110 VAC, 250 VAC maximum |
| OutputPulse
|
pulse length (0.5 to 20 sec) or pulse frequency (60 to 120 pulses per min) |
| OutputCurrent
|
isolated 0-20.4-20 mA, with 20% to 100% full-scale scalable boundaries, 600 maximum load |
| Hysteresis
|
0.1 to 1.0 mg/l; 1 to 10% |
| Relay delay
|
selectable, 0 to 2000 seconds |
| Input impedance
|
10^12 M
|
| Electrical isolation
|
yes, galvanically |
| Salinity correction
|
manual input from 0.0 to 50.0 ppt; automatic correction |
| Barometric pressure correction
|
manual input from 0.0 to 50.0 ppt; automatic correction |
| Temperature Sensor
|
100 Pt RTD |
| Temperature Compensation
|
Automatic or manual from -9.9 to 125°C |
| Password protection
|
two level password |
| Display
|
dual line LCD, 4-digit upper and 3 1/2 digit lower |
| Operating Temperature
|
-10 to 60°C |
| Housing
|
1/4 DIN |
| Dimensions
|
3.8" x 3.8" x 6.9" (9.6 x 9.6 x 17.5 cm) |
| Shipping weight
|
2.5 lbs (1.2 kg) |