| Measures
|
pH, mV, °C |
| pHRange
|
0.00 to 14.00 pH |
| pHResolution
|
0.01 pH |
| pHAccuracy
|
±0.01 pH |
| pHCalibration
|
one or two points (select from pH 1.00, 4.01, 6.86, 7.00, 9.00, 9.18, or 10.00) |
| mVRange
|
±1000 mV |
| mVResolution
|
1 mV |
| mVAccuracy
|
±1 mV |
| mVCalibration
|
mV offset |
| TemperatureRange
|
-9.9 to 125°C |
| TemperatureResolution
|
0.1°C |
| TemperatureAccuracy
|
±0.1°C |
| TemperatureCalibration
|
offset up to ±5°C |
| Control type
|
on/off |
| Number of inputs
|
one |
| Number of set points
|
two (high and low) |
| Separate alarm relay
|
no |
| OutputControl
|
2 SPDT relays, 6 A at 110 VAC, 250 VAC maximum |
| OutputCurrent
|
isolated 4-20 mA, 600 W maximum load |
| Hysteresis
|
0.1 to 1.0 pH, 10 to 100 mV, or 1 to 10% |
| Relay delay
|
selectable, 0 to 2000 seconds |
| Input impedance
|
10^6 MW
|
| Electrical isolation
|
yes, galvanically |
| Temperature Sensor
|
100 W or 1000 W Pt RTD |
| Temperature Compensation
|
Automatic or manual from -9.9 to 125°C |
| Password protection
|
two level password |
| Display
|
dual line LCD, 4-digit upper and 3 1/2 digit lower |
| Operating Temperature
|
-10 to 60°C |
| Housing
|
1/4 DIN |
| Dimensions
|
3.8" x 3.8" x 6.9" (9.6 x 9.6 x 17.5 cm) |
| Shipping weight
|
2.5 lbs (1.2 kg) |