HygroFlex M
Software


  Rotronic HygroFlex M Software
  Airport Weather Stations - Meteorological Towers - Rooftop Installations

HW3 is a software package designed for the HygroClip product line. HW3 features data acquisition, recording, graphing and alarm functions, calculation of various humidity parameters (both on-line and off-line) as well as instrument configuration and probe calibration. HW3 also offers a comprehensive solution for the networking of up to 32 HygroFlex M3 transmitters.


Network Display Form

• On-line presentation of the data in tabular form
• Up to 32 transmitters
• Up to 64 humidity and temperature probes
• Up to 192 measured and calculated values
• Live graph of up to 8 measured or calculated values
  with or without data logging
Data-logging

• Log interval adjustable from a few seconds to several hours
• Up to 32 instruments can be simultaneously recorded
• Individual log file per instrument or all instruments on the network
  in a single log file (requires Microsoft's Excel)
• Size of log file can be limited based on the number of samples or a new file
  can be automatically started each hour, day or month
• Display or print-out of the data in tabular format
• File data can be transferred to HW3 graphic module or to Microsoft's Excel
Alarm function

• Monitors relative humidity and / or temperature
• Can be set for individually for each instrument and probe on the network
• Acoustic and or visual signal directly on PC
• Remote alarm by email (requires Microsoft's Outlook)
Graphs

• On-line (live graph) or off-line (based on a log file)
• Can display calculated values: dew-point, mixing, ratio etc.
• Up to 16 traces simultaneously
Instrument Display Form

• On-line presentation of the data for a single transmitter
  (number of probes depends on the model and its options)
• Selection of the calculated parameter
• Access to data logging, graph and alarm functions
• Access to instrument configuration and probe calibration


Probe Calibration (HygroClip probes)

• Humidity: up to 4 points, Temperature: 2 points
• 1-point adjustment function (offset)
• Automatic temperature correction of the ROTRONIC humidity standards
• Certificate can be generated


System Requirements


PC




Pentium 233 MHz or better
Windows 95, 98, Me or NT, 2000
Memory 32 MB or better
Disk Space 20 MB
Monitor VGA or Super VGA
Resolution 1024 x 768 recommended (800 x 600 minimum)
High Color 16 Bit or better (256 color minimum)
Connection RS232 (COM 1 - 4)
Software HW3 (includes Browser, Adobe Acrobat Reader)
Technical Data
Humidity sensor ROTRONIC Hygromer® C94 (capacitive)
Temperature sensor RTD Pt100 Ohm, 1/3 DIN
Sensor protection Wire mesh filter (standard) / foam filter (optional)
Measuring range 0...100%RH and -40...85°C ( -40...185°F)
Operating limits same as measuring range
Accuracy at 23°C / 73°F ±1.5%RH and ±0.2°C / 0.4°F (when calibrated for HygroFlex M)
Repeatability better than 0.5%RH and 0.1°C/°F
Analog output signals
(minimum load > 10 k Ohm)
0...100%RH = 0...1V
- 40...85°C = 0...1.25V (-40...60°C = 0...1V)
Digital output/input signal 1-wire, HygroClip format, 2-way communication
Analog signal resolution 0.02%RH and 0.06°C
Digital signal resolution 0.004%RH and 0.004°C
Sampling time < 0.7 sec
Maximum cable length 5 m /15 ft for direct output
up to 100m / 325 ft with signal amplifier
Connection type bayonet cap on mounting connector (see accessories)
Protection grade IP65 (probe to connector)
Supply Voltage 3.5...50VDC
Current Consumption < 4mA
Minimum excitation time < 4 sec
EMC compatibility (CE) EN-50081-2, EN50082-2
Material Polycarbonate, white
Dimensions length: 100mm (3.94"), diameter: 15mm (0.59")
Weight 14 g / 0.5 oz
Probe Calibration

For scheduled maintenance, you may to want to rotate the probes (as opposed to the
complete instrument) between the measurement site and a calibration laboratory. The
HygroClip product line is designed to make this easy.

After removing the probe from the HygroFlex M transmitter, the following alternatives are available to calibrate the probe:

  1. Calibrate the probe using a HygroPalm 3 indicator
  2. Calibrate the probe using a PC on which the HW3 software has been installed (requires the MOK-03-WIN calibration cable to connect the probe to the PC).
  HW3 Software: software based probe calibration is both simple and precise: there are no hard-to-reach, hard-to-adjust potentiometers. HW3 offers a high level of flexibility in selecting the calibration reference, giving the user complete control over the calibration process. HW3 features the following calibration modes:

• Humidity: 2, 3 or 4 points
• Temperature: 2 points
• 1-point adjustment function (offset) for both humidity and temperature

When humidity calibration is done using the Rotronic humidity standards, HW3 automatically corrects the value of the standard to reflect the temperature of calibration (see note). HW3 also offers the possibility of printing a certificate of calibration.

Note: the value of the standards does not depend on barometric pressure (altitude).

 

Certified Humidity Standards:
if you do not have access to a humidity generator, the Rotronic humidity standards offer a reliable and convenient way to generate a reference humidity condition during probe calibration. The Rotronic standards are shipped with an internationally accepted certificate, meeting in full the requirements of quality systems such as ISO 9000. The standards are available in boxes of 5 glass ampoules of the same value, and can be stored indefinitely. Standards in the range of 5 to 95 %RH are non-saturated aqueous salt solutions. The 0 %RH humidity standard is made of small granules of a highly porous ceramic. A Material Safety Data Sheet is available for each standard.

 

Calibration Devices:
a calibration device is a small airtight container that fits on the probe and seals around the humidity sensor. During calibration, a known reference humidity is produced inside the calibration device by means of a humidity standard.

Four calibration devices are available for the HygroClip probes:
ER-15 for 15mm diameter probes
ER-05 for 5mm diameter probes
EM-15 for type ‘IE’ probes
ERV-15 for 15mm diameter probes (wall mounting)
Probe Calibration Accessories
MOK-03-WIN Calibration cable for HygroClip

HW3 Calibration software on CD ROM
ER-15 Calibration device for HygroClip
EAXX-SCS Certified humidity standards (5 per box)
XX = 00. 05, 10, 20, 35, 50, 65, 80, 95%RH