Signal
Conditioner
Validyne Model CD15
Description
Features
Specifications
Instruction Manual
Typical CD15 Connections
Signal
Conditioner Description:
The CD15 Signal Conditioner is
a basic carrier demodulator which provides a +/-10 Vdc analog
output over the full pressure scale of any Validyne variable reluctance
pressure transducer.
The CD15 Signal Conditioner supplies a 5 Khz excitation to the
transducer and provides zero balance and span adjustments mounted
on ten-turn dials. The analog output signal is available from
the front panel on two binding posts.
The CD15 Signal Conditioner also includes a low pass output filter
which is flat to 1000 Hz.
The CD15 Signal Conditioner normally operates on 115 VAC power
but models are available for 220 VAC and 12 Vdc. Output signal
options include 4-20 mA, +/-1 Vdc and square root extraction.
Signal
Conditioner Features:
- Provides +/-10 Vdc Output for
any VR Transducer
- Compact for Bench Top
- 110 VAC Powered
- Easy to Use
Signal
Conditioner Specifications:
Power:
95-125 VAC, 50-400 Hz (Std)
190-250 VAC, 50-400 Hz (Optional)
10.5 - 14.5 Vdc (Optional)
Output Signals:
+/-10 Vdc (Std)
+/-1 Vdc
4-20 mA
Square Root Outputs Available
Gain Range:
+/-15 to +/-75 mV/V Input
for +/-10 Vdc FS Output
Zero & Span Adjustments
Low Pass Filter:
1000 Hz
Size & Weight:
5.6" x 5.5" x 3"
3.5 Lbs
Typical
Connections

CD15 Carrier Demodulator
Instruction Manual
Description
The CD15 Signal Conditioner is
a sine wave carrier demodulator designed to operate with variable
reluctance transducers to provide a DC output signal for dynamic
as well as steady state pressure measurements. The instrument
furnishes regulated carrier excitation of 5 Vrms, 5 Khz sine wave
to the two inductance arms of the pressure transducer. The sensor
output is demodulated, amplified and filtered to produce a +/-10
Vdc analog output for the full scale range of the pressure transducer.
The low impedance sine wave excitation allows the CD15 Signal
Conditioner to be located several hundred feet away from the pressure
transducer.
Zero and span controls are provided
to adjust for unbalance in the system and to set the output gain
of the demodulator to match the transducer sensitivity. An external
balance pin is provided for additional permanent zero offset capability.
All circuits are operated from a built-in regulated +/-15 Vdc
power supply. Power is from 110 VAC , 60 Hz line (standard version).
Controls and Connections
Two ten-turn potentiometers with
locking dials are located on the front panel . One control is
for setting the zero adjust and the other for setting span during
calibration.
Two terminal posts, one red (+)
and one black (-) are located on the front panel for the DC signal
output.
A toggle on/off power switch is
located on the front panel for energizing the instrument and applying
power to the transducer connected to the CD15. The unit also contains
an LED pilot light which comes on when power is applied.
A 115 VAC, 60 Hz power cable (standard)
plugs into any standard AC outlet. A transducer connector is mounted
on the rear panel and connects to the cable from the variable
reluctance pressure sensor. This connector may be a six-pin Bendix
type, PT02A-10-6P or a 4-pin Cannon type, WK4-32S.
Options
A complete breakdown of options
as defined by the model number is given.
Technical Characteristics
Input Sensitivity: 15 mV/V,
minimum
Transducer Excitation: 5 Vrms, 5 Khz Sine Wave
Zero Adjustment Range: +/-10 mV/V minimum, at 10% of span
Span Adjustment Range: 0 to 100% of input
Output Voltage: +/-10 Vdc @ 10 mA, short circuit proof
Output Impedance: 10 Ohms, nominal
Frequency Response: DC to 1000 Hz, flat +/-10% (see model
options)
Stability: +/-0.1%/30 days
Ripple: < 10 mV peak-to-peak
Zero Shift: +/-0.005%/deg
F, average
Operating Temperature Range: 0 to 185F
Power Requirements: 105 to 125 VAC, 50-400 Hz, 5 W
Weight: 3.5 lbs
Dimensions: See Outline Drawing
Installation
The CD15 Signal Conditioner can
be operated on a bench or attached to a panel for rack-mounting.
Transducers are connected to the CD15 by means of the connector
on the rear panel. Cable shields should be connected to the shells
of the mating connectors. In cases where the transducer body is
grounded, the shield connection at the transducer end should be
left open in order to eliminate noise from possible ground currents.
See also Cabling Considerations.
The only electrical connections
required are plugging the instrument into the appropriate power
source, connection of the transducer cable and connection of the
output binding posts to the receiving instrument.
Transducer Calibration
1. Connect the transducer cable
to the connection on the back panel. Note: The same transducer
cable should be used during calibration as in actual use as variations
in cable lengths can affect the span settings.
2. Connect the transducer to a
pressure source, and bleed the system .
3. Connect a voltmeter to the
red (+) and black (-) output posts on the CD15 front panel. (See
Diagram)
4. Reduce the input pressure to
0, and adjust the front panel zero control for an output of 0
Vdc.
5. Apply full scale pressure to
the transducer and adjust the span control for +10 Vdc ouput.
Note that any full scale output voltage may be set, from 0 to
+10 V.
6. Reduce the pressure to 0 and
re-chck the output, adjusting the zero control as necessary. Engage
the zero and span adjustment locks.
7. The transducer, cable and CD15
Signal Conditioner are now calibrated together as a system. Record
the zero and span dial settings. When multiple transducers are
used with the same CD15, record zero and span settings separately
as each transducer will have a unique pair of settings.
Transducer Connector Pin Connections
For Models With Six-pin Bendix
PT Style Connector:
Pin A - 5 Vrms, 5 Khz Excitation
Pin B- Transducer Input Signal
Pin C - Extrernal Balance Pin
Pin D - 5 Vrms, 5 Khz Excitation
For Models With Four-pin Cannon
WK4 Style Connector:
Pin 1 - 5 Vrms, 5 Khz Excitation
Pin 2 - Transducer Input Signal
Pin 3 - Extrernal Balance Pin
Pin 4 - 5 Vrms, 5 Khz Excitation
Model Numbers
Outline Drawing
Cabling Considerations
for Variable Reluctance Pressure Sensors
Variable Reluctance pressure sensors
operate on an inductive sensing principle that requires and AC
excitation and returns an AC signal. Long cabling runs from the
sensor to the carrier demodulator are possible without compromise
in performance if the proper cable configuration is observed.
This application note describes recommended cable geometries for
variable reluctance sensors such as models DP15, DP45 or DP103.
The purpose of using the correct
cabling configuration is to minimize cross-talk between the carrier
wires and the signal wires. If the distance between the sensor
and carrier demodulator is just a few feet, no precautions are
necessary: even loose pigtail leads will be sufficient. If the
distance between the demodulator and sensor is over ten feet,
it is best to select one of the cable geometries shown in the
diagram below.
The optimal cable type consists
of two pairs of separately shielded wires. This is the standard
used for all Validyne cable assemblies. Belden #8434 cable has
an inner-shielded pair (red and black) and an outer pair (green
and white). The two inner-shielded wires are connected together
to form the signal line, and this is wired to pin 2 or pin B of
the cable connector. The green wire is the positive phase of the
carrier, wired to pin A or pin 1, and the white wire is the negative
carrier wired to pin 4 or pin D. The signal line is shielded from
the excitation wires so that inductive cross-talk to the pressure
signal is minimized.
The shields of the cable are generally
connected to the connector shell, which, in turn, is mechanically
connected to the demodulator and transducer housings. In cases
where a difference in ground potential is expected between the
plumbing and electronics, the shield should be disconnected at
the transducer.
The following diagram shows various
possible cabling configurations:

Typical Setup
|