Dual,Thermocouple,Input,Thermometer,J,T,E,K,Barnant

  Dual Thermocouple Input Thermometer J - T - E - K

  Index  New Products   Contact Us   Tech Support   Site Search
Products
 Calibrators
 Controllers
 Counters/Timers
 Hygrometers
 Indicators - Digital
 Load Handling Devices
 Manufacturers
 Recorders
 Rupture Disks
 RTD's
 Sensors
 Temperature Products
 Test Equipment
 Thermocouples
 Transducers
 Transmitters

Quality Temperature and Recording Instrumentation
...at the lowest prices anywhere!

 












Dual Input Thermometer J- T- E- K

  • Accepts up to 2 thermocouple inputs and either J, K, T or E thermocouples
  • Manually store up to 25 readings in non-volatile memory
  • Two point field calibration; offset and slope for improved system accuracy
  • T1/T2 differential
  • Accuracy: ±0.7°F (±0.4°C) above -238.0°F (-150.0°C); plus ±0.1% of reading
  • UL & cUL rated Intrinsically safe-Certified for Class I, Division 1; Groups A, B, C, and D hazardous (classified) locations
  • Sealed Silicone rubber keypad
  • Rated IP 54 for splash & dust resistance
  • ABS/polcarbonate impact resistant case
  • Fold out stand for bench top operation
  • Meets CE requirements for electromagnetic compatibility
  • Bright, easy to read LCD
  • 3 year warranty
  • Manufactured in U.S.A.


FEATURES:

MIN, MAX, HOLD, °C OR °F, MULTI-DATA LCD

This thermometer offers you numerous features; choose from J. K, T or E thermocouple curves, dual probe capability, selectable resolution and 12 button keypad for simplified set up.

Display Minimum, Maximum and differential readings simultaneously using one or two probes on selectable, easy-to-read, multi-data LCD. Two point field calibration feature for offset and slope. Allows user to calibrate the probe to known temperatures. Meter automatically stores and adjusts for inherent probe error enabling unparalleled system accuracy. Calibration lock out protects NIST traceable factory settings and prevents tampering with field calibration values.

ORDERING INFORMATION
600-1040 Dual J-T-E-K Thermometer
Cat. no. Type Range Accuracy Resolution Functions Display
600-1040 J -328 to 1832°F
(-200 to 1000°C)
above -150°C (-238°F)

±(0.1% of reading and ±0.4°C (0.7°F)

below -150°C (-238°F)

±(0.25% of reading and ±1°C (2°F)

0.1°:
<= 999.9°
and 1°: > 999.9°
On/off, hold, calibration, °C/°F, J-T-E-K, maximum, minimum, differential, store, Three 4-digit LCD. Displays: 0.4"H and 0.2"H
K -418 to 2501°F (-250 to 1372°C)
T -418 to 752°F (-250 to 400°C)
E -418 to 1832°F (-250 to 1000°C)



Dual Thermocouple Input Thermometer J - T - E - K from Barnant
Cyronix Instrumentation & Controls
12 Upjohn Road, Unit 5
Don Mills, Ontario, M3B 2V9
Tel: 416 444-6614 Fax: 416 444-0353
Toll Free: 1 888-CYRONIX
 