Temperature/Humidity
Logger
- Measure temperature and
humidity with one instrument
- Accepts two type J, K,or
T thermocouple probes
- Standby mode conserves
batteries while logging data
- Optional software for bidirectional
RS-232 communications
- Stores 850 pairs of measurements
Use the benchtop
Temperature/Humidity
Logger almost
anywhere temperature and humidity needs to be monitored
- laboratories, museums, art galleries, warehouses,
protected outdoor environments, and more! The logger
operates in three modes: two-channel temperature,
humidity/temperature, and dew point/temperature.
This compact, portable Temperature/Humidity
Logger is packed with
features. Nonvolatile memory stores setup conditions
(interval time or total logging time), thermocouple
probe type, logged/stored parameters, and peak/valley
values for each mode. Serial input/output allows
for printer and computer compatibility, Optional
data analysis software and serial printer available.
Measure and display either
differential (DT) temperature or dual temperature
simultaneously in the two-channel temperature mode.
The temperature
/ humidity logger accepts
one or two type J, K, or T thermocouple probes with
ANSI miniconnectors. Features in this mode include
autocalibration of probe, selectable °F or °C
scale, selectable 0.1%/1° resolution up to 999.9°
(1° above 999.9°), dual analog temperature
output, and accuracy of ±0.1% of reading.
Measure and display both relative
humidity and temperature (RTD) in the humidity/temperature
mode. Use one of two interchangeable relative humidity/temperature
probes. Wide-range probe 637-0050 has a compensated
RH range from 0 to 100%. General-purpose probe 637-0052
has a compensated RH range from 5.0 to 95.0%. Other
features in this mode include: automatic field calibration,
a humidity resolution of 0.1% and 1% and analog
humidity/RTD outputs.
In the dew point/temperature
mode, temperature
/ humidity logger calculates
the dew point from the relative humidity and temperature
data, and displays it and the temperature (RTD)
on the LCD. A dew point range of -94° to 158°F
(-70° to 70°C) is available with wide-range
humidity probe 637-0050, a dew point range of -58°
to 122°F (-50° to 50°C) is available
with general-purpose humidity probe 637-0052.
The Temperature/Humidity
Logger is housed in
a molded ABS plastic housing and has a membrane
keypad. Dual LCD shows scale and mode selected.
Two AA batteries (included) provide 50 hours of
continuous operation in standby mode. Purchase optional
battery pack for up to 300 hours of continuous use.
(We recommend using the optional AC wall mount power
supply 691-9050 where line power is available).
Order the L-shaped mounting bracket to attach either
RH probe to the logger. See below to order software,
logger kits, probes, and accessories.
Temperature/Humidity
Logger Specifications:
WITH THERMOCOUPLE PROBES:
Temperature ranges:
J: -310 to 1832°F (-190
to 1000°C)
K: -418 to 2507°F (-250
to 1375°C)
T: -418 to 752°F (-250
to 400°C)
Temperature resolution: 0.1/1° F
or °C (switchable); autoranging to 1°
above 999.9°.
Temperature accuracy: ±0.1% of reading
or ±0.7°F (±0.4°C) above -238°F
(-150°C) ±2°F (±1°C) below
-238°F (-150°C)
Differential (DT) accuracy: ±0.5°F
(±0.3°C)
Probes: one or two type J, K, or T thermocouple
probes (sold separately)
|
WITH RELATIVE HUMIDITY/TEMPERATURE
PROBE:
RH ranges:
637-0050 0.0 to 100.0% RH
within -22 to 158°F (-30 to 70°C) temperature
range
637-0052 5.0 to 95.0% RH
within 32 to 122°F (0 to 50°C) temperature
range
RH resolution: 0.1% / 1% (selectable)
RH accuracy:
637-0050 ±2% RH from
10.0 to 96.0% RH; ±3% outside these limits
637-0052 ±RH from 10.0
to 90.0% RH; ±3% outside these limits
Temperature range: -32 to 212°F (0
to 100°C)
Temperature accuracy: ±2.7°F
(±1.5°C)
|
GENERAL SPECIFICATIONS:
Display: dual LCD - one 4-digit, 3/8"H
and one 5-character alphanumeric, 3/16"H
Output: 0.0 to 1.0V analog output of
temperature and humidity; bidirectional RS-232-C
Memory: Stores 850 pairs of readings
for recall
Power: two AA batteries (included), optional
battery pack, or optional AC adapter
Battery life: up to 50 hours of continuous
operation in standby mode (with AA batteries)
or up to 300 hours (with optional battery pack)
Low battery indication: yes
Dimensions: 6 1/2"Wx3 1/2"Hx7
3/4"D
Shipping weight: 3 lbs (1.4 kg)
|
ORDERING
INFORMATION
691-9000 Barnant Temperature/Humidity Logger
|
Temperature/Humidity
Logger Kit
Keep your logger and accessories together with
this handy kit. Unit includes temperature/humidity
logger, one general-purpose relative humidity/temperature
probe (637-0052), one L-shaped
humidity probe mounting bracket, one 10-ft humidity
probe extension cable (637-0055),
two type K thermocouples (8516-82),
pack of four AA batteries, 115 VAC wall mount
power supply, instruction manual, data analysis
software, and deluxe carrying case (19"W
x 14 1/4"H x 5 1/4"D)
691-9020 Temperature/Humidity
Logger kit; 115 VAC, 50/60 Hz. Shpg wt. 7 lbs
(3.5 kg)
|
 |
 |
Probes
(measure 12 mm dia with 5"L handle)
|
| Cat
no. |
Description |
Shpg
wt. |
| 637-0050 |
Wide-range
RH/temp, 12"L |
1
lb (0.5 kg) |
| 637-0052 |
General-purpose
RH/temp, 3 3/4"L |
1
lb (0.5 kg) |
Cables (for sending
data to a printer or computer)
|
| Cat
no. |
Description |
Shpg
wt. |
| 692-8050 |
Cable
to DB25M serial printer, 6 ft |
1
lb (0.5 kg) |
| 692-8052 |
Cable
to DB9F serial computer port, 6 ft |
1
lb (0.5 kg) |
| 692-8054 |
Cable
to DB25F serial computer port, 6 ft |
1
lb (0.5 kg) |
| Cat
no. |
Description |
Shpg
wt. |
| 691-9060 |
L-shaped
wall/meter probe mount bracket |
1
lb (0.5 kg) |
| 691-9080 |
10-ft
humidity probe extension cable |
1
lb (0.5 kg) |
| 3335-00 |
Humidity
meter calibration kit* |
5
lbs (2.5 kg) |
| Data
Analysis Software
Analyze and display
Temperature and Humidity data on your WINDOWS®
95/98/NT4. You can use data from previously
stored file or from the temperature/humidity
logger. Calculate and display additional data
such as absolute humidity, water vapor weight
and pressure, humidity ratios, specific volume
and enthalpy. Create charts and graphs, download
and store real-time temperature and humidity
data, and configure and check logger setup.
Software on CD-ROM with online documentation.
|
 |
|