The system is supplied with a TP1 scissor action pump or the LTP1 low-pressure pump with fine control. Both pumps incorporate adjustable over pressure protection, to avoid over pressuring units under test. The clear LCD display has a back light for easy use in poor light situations and is small and portable. The PC6-AV Pressure Calibrator has an accuracy (FS) better than 0.025% and resolution (FS) better than 0.01%. Pressure units can be recorded on the PC6-AV in mbar, inHg, knots, kPA, psi, atm, in H2O, mH2O, mmH2O, bar, mmHg, unit1, unit2 with altitudes measured in feet or meters with a range of -3,000 ft. to 105,000 ft. Non-corrosive gases are the preferred pressure media (contact Cyronix for other media), and the pressure connections are 1/4 BSP, or 1/4 NPT male (8mm O/D hose connector). Temperature effects are 0.001% of reading/ºC between 10-40ºC, and data capacity is 175 records into a maximum of 20 files, 8K memory with an option to increase that capacity by an additional 8K. RS232 communication is included in the design and the optional SI-CAL software will present downloaded data in various graphic forms such as gauges or a scrolling bar and remotely control the PC6-AV Pressure Calibrator via the RS232 cable linked to a PC. Data storage is compatible with most analysis, database and word processing programs. |