| OPEN CHANNEL FLOW MONITOR
SPECIFICATIONS
SCOPE: This specification
covers an ultrasonic, echo-ranging type open channel flow
meter as manufactured by Greyline Instruments, Massena,
New York / Long Sault, Ontario. This instrument shall
provide for indicating, transmitting, totalizing and data
logging of the flow rate through a flume, weir, or other
primary measuring device.
A. GENERAL
- Open Channel Flow Monitor
to consist of an non-contacting ultrasonic sensor, connecting
cable, and a remote enclosure with indicating, transmitting
and controlling electronics.
- Measurement accuracy
shall be ±0.25% of Range or 2 mm (0.08"), whichever
is greater, and shall be automatically temperature compensated.
- Sensor cable length
shall be as required by installation, not to exceed
500' (152 m).
- System shall have no
moving parts and shall not contact the water being measured.
- Shall include PC software
program disk to determine open channel flow calibration
values for non-standard flumes and weirs.
B. SENSING ELEMENT
- Sensor shall withstand
accidental submersion to 10 ft (3 m) maximum.
- Sensor minimum range
shall be 8" (203.2 mm) and maximum range shall
be 15 ft (4.57 m).
- Sensor shall be constructed
of PVC.
- Sensor operating temperature
shall be from -40°F to 150°F (-40°C to 65°C).
- Sensor shall include
integral temperature sensor. Temperature sensors requiring
separate mounting and wire runs shall not be accepted.
C. SENSOR CONNECTING
CABLE
- Provide RG62AU coaxial
cable 25' (7.6m) continuous length, with waterproof,
potted bond to the Sensor head.
- Extended sensor cable
shall be RG62AU coaxial to a maximum of 500' (152m).
Cable shall be spliced with screw terminal connections
in manufacturer's recommended steel NEMA4 Junction Box.
- Level and temperature
signals shall be conducted on one single coaxial cable.
Separate or multiple-conductor cables shall not be accepted.
- Sensor cable shall be
installed in grounded metal conduit.
D. TRANSMITTER
- The transmitter shall
provide for field calibration to common primary metering
devices, plus allow entry of "K" and "n"
values for calibration to nonstandard flumes, weirs
or open channels. Systems
requiring calibration by Parameter codes, BCD switches
or external calibrators shall not be accepted.
- Calibration data shall
be password protected and permanently stored through
power interruptions for a minimum of 12 months.
- Field calibration shall
allow selection and automatic conversion of measurement
units, measurement span, high/low flow alarm relay and
flow proportional relay pulse rates.
- Transmitter shall permit
field programmable damping to smooth output in turbulent
flow conditions, and programmable rejection time to
disregard false signals from rain, wind, floating objects
and spurious echoes.
- Transmitter operating
temperature shall be from -5° to 140°F (-20° to 60°C).
Transmitter shall contain a thermostat-controlled enclosure
heater for condensation protection below 30°F (-1°C).
- Transmitter shall have
an isolated 4-20mA output rated for 1000 ohm maximum
load with menu-selectable 0-5VDC alternative.
- Have a built-in 2
million point Data Logger with USB output to flash
drive or mass storage device. Include Windows software
for data log analysis, graphing and export.
- Provide two relay contacts
rated 5 amp SPDT programmable for single set point alarms,
dual set point pump control, pump alternation, temperature
alarm, flow totalizer pulse and/or echo loss alarm.
- Provide a white, backlit
matrix LCD display indicating flow rate, level, totalizer
and relay states in user-selected engineering units
and with language selection including English, French
and Spanish.
- Transmitter display
indicating level or flow rate, units of calibration,
totalizer and relay states shall be visible without
opening cover.
- Transmitter shall be
housed in a wall-mount, watertight NEMA4X (IP66) enclosure
with hinged, clear cover. Mounting hardware shall be
included.
- Transmitter electronics
shall be surge protected on AC power input, sensor and
4-20mA/0-5V outputs.
- Transmitter power input
shall be 100-240VAC 50-60Hz with power consumption of
4.0 Watts or less.
- The transmitter
shall permit plug-in field installation and auto-detection
of optional accessories including additional control
relays.
E. OPTIONAL FEATURES
FOR INSERTION AS REQUIRED
- Have power input of
9-32VDC and power consumption of less than 4.0 Watts.
- Sensor cable shall be
50 ft. (15 m) continuous length with potted bond to
the Sensor head.
- Sensor cable shall be
100 ft. (30 m) continuous length with potted bond to
the Sensor head.
- Separate length extended
sensor coaxial cable shall be extended to a total maximum
length of 500 ft (152m). Cable shall be spliced with
screw terminal connections in manufacturers recommended
steel NEMA4 Junction Box.
- Sensor, connecting cable
and junction boxes shall be rated intrinsically safe
to Class I,II,III, Div. I,II, Groups C,D,E,F,G.
- Have a thermostat controlled
enclosure heater for Transmitter operation at temperatures
below freezing.
- Have manufacturer's
recommended enclosure sunscreen to permit Transmitter
mounting in direct sunlight.
- Have manufacturer's
recommended Sensor sunscreen to permit temperature-compensated
Sensor mounting in direct sunlight.
- Have manufacturer's
recommended Panel Mount Flange assembly for enclosure
installation.
- Have 4 additional (6
total) control relays, rated 5 amp SPDT and programmable
for single set point alarms, flow proportionate pulse,
temperature alarm and/or echo loss alarm.
F. MANUFACTURER
Open
Channel Flow Monitor shall be Model OCF 5.0 as manufactured
by Greyline Instruments Inc. and shall be warranted against
defects in materials and workmanship for one year.
|