| LEVEL & FLOW MONITOR
SPECIFICATIONS
SCOPE: This specification
covers a non-contacting ultrasonic Level/Flow Monitor
as manufactured by Greyline Instruments, Massena New York
/ Long Sault, Ontario. This instrument shall provide for
indicating, transmitting and control of the material level
in a vessel or proximity of a target to the instrument
sensor, and indicating, transmitting, and totalizing of
the flow rate through a flume, weir, or other primary
measuring device.
A. GENERAL
- Level/Flow Monitor to
consist of an non-contacting ultrasonic sensor, connecting
cable, and a remote enclosure with indicating, transmitting
and controlling electronics.
- Measurement accuracy
shall be ±0.25% of Range or 2 mm (0.08"), whichever
is greater, and shall be automatically temperature compensated.
- Sensor cable length
shall be as required by installation, not to exceed
500' (152 m).
- System shall have no
moving parts and shall not contact the material being
measured.
- Shall include PC software
program disk to determine open channel flow calibration
values for non-standard flumes and weirs.
B. SENSING ELEMENT
- Sensor shall be constructed
of PVC and Teflon.
- The sensor shall have
a minimum deadband or blanking of 12" (305 mm)
and a maximum range of 32 ft. (10 m), and have an operating
frequency of 42 kHz with an ultrasonic beam angle of
8 Degrees.
- Sensor shall withstand
accidental submersion to 20 psi.
- Sensor operating temperature
shall be from -40°F to 150°F (-40°C to 65°C).
- Sensor shall include
integral temperature sensor. Temperature sensors requiring
separate mounting and wire runs shall not be accepted.
C. SENSOR CONNECTING
CABLE
- Provide RG62AU coaxial
cable 25' (7.6m) continuous length, with waterproof,
potted bond to the Sensor head.
- Extended sensor cable
shall be RG62AU coaxial to a maximum of 500' (152m).
Cable shall be spliced with screw terminal connections
in manufacturer's recommended steel NEMA4 Junction Box.
- Level and temperature
signals shall be conducted on one single coaxial cable.
Separate or multiple-conductor cables shall not be accepted.
- Sensor cable shall be
installed in grounded metal conduit.
D. TRANSMITTER
- The transmitter shall
provide for field-calibration via built-in 5-key calibration
system with menu selection of parameters. Systems requiring
calibration by Parameter codes, BCD switches or external
calibrators shall not be accepted.
- Calibration data shall
be password protected and permanently stored through
power interruptions for a minimum of 12 months.
- Field calibration shall
allow selection and automatic conversion of measurement
units, measurement span and control relays.
- The transmitter shall
provide for field calibration in user-selected Range,
Level or Open Channel Flow modes. Flow mode shall allow
calibration to common primary metering devices, plus
allow entry of calibration formula for non-standard
flumes, weirs or open channels.
- Transmitter shall permit
field programmable damping to smooth output with turbulent
level.
- Transmitter operating
temperature shall be from -5° to 140°F (-20° to 60°C).
Transmitter shall contain a thermostat-controlled enclosure
heater for condensation protection below 30°F (-1°C).
- Transmitter shall have
an isolated 4-20mA output rated for 1000 ohm maximum
load with menu-selectable 0-5VDC alternative.
- Provide two relay contacts
rated 5 amp SPDT programmable for single set point alarms,
dual set point pump control, pump alternation, temperature
alarm, flow totalizer pulse and/or echo loss alarm.
- Provide a white, backlit
matrix LCD display indicating flow rate, level, velocity,
totalizer and relay states in user-selected engineering
units.
- Transmitter display
indicating level or flow rate, units of calibration,
totalizer and relay states shall be visible without
opening cover.
- Transmitter shall be
housed in a wall-mount, watertight NEMA4X (IP66) enclosure
with hinged, clear cover. Mounting hardware shall be
included.
- Transmitter electronics
shall be surge protected on AC power input, sensor and
4-20mA outputs.
- Transmitter power input
shall be 100-240VAC 50-60Hz with power consumption of
3.5 Watts or less.
- The transmitter
shall permit plug-in field installation and auto-detection
of optional accessories including data logger and additional
control relays.
E. OPTIONAL FEATURES
FOR INSERTION AS REQUIRED
- Have 4 additional (6
total) control relays, rated 5 amp SPDT and programmable
for single set point alarms, dual set point pump control,
pump alternation, temperature alarm, flow totalizer
pulse and/or echo loss alarm.
- Sensor, connecting cable
and junction boxes shall be rated intrinsically safe
to Class I,II,III, Div. I,II, Groups C,D,E,F,G.
- Have a thermostat controlled
enclosure heater for Transmitter operation at temperatures
below freezing.
- Have an all-Teflon sensor
Model PZ32TE rated intrinsically safe when connected
through an approved safety barrier, and rated for operation
from -40°F to 170°F (-40°C to 76°C).
- Have a PVC and Teflon
sensor Model PZ52T rated for measurement range on liquids
up to 50 ft. (15.6 m).
- Have a Teflon-faced
flange mount sensor model PZ34TxF rated for 32 ft (10
m) measurement range. Specify flange size of 4”, 6”
or 8”.
- Have manufacturer's
recommended enclosure sunscreen to permit Transmitter
mounting in direct sunlight.
- Have manufacturer's
recommended Sensor sunscreen to permit temperature-compensated
Sensor mounting in direct sunlight.
- Have manufacturer's
recommended Panel Mount Flange assembly for enclosure
installation.
- Have a built-in 2
million point Data Logger with USB output to flash
drive or mass storage device. Include Windows software.
- Have power input of
9-32VDC and power consumption of less than 3.5 Watts.
F. MANUFACTURER
Level/Flow
Monitor shall be Model SLT 5.0 as manufactured by Greyline
Instruments Inc. and shall be warranted against defects
in materials and workmanship for one year.
|