The
DataChart 3600 Series Paperless Chart Recorders offer
the lowest per channel cost for a 3/8 din recorder. They have
a total of 15 channels and will take up to 12 direct universal
inputs including: DC Voltage, DC Current, Thermocouples and
RTD's.
DataChart 3600 Series Paperless Chart Recorders comes
standard with RS232 communications and an extraordinarily powerful
math package. 21 bit A/D converters yield high measurement specifications
of +/- .05%.
Information can be stored on standard 3 1/2" disks or PCMCIA
flash cards up to 220 Megabytes! The number of channels and
the storage frequency will help you decide how much storage
you need.
The
DataChart 3000/4000 Paperless Chart Recorders Series was designed
from the ground up with flexibility in mind. All options
are plug-n-play and field installable, so expanding the number
of input channels or adding relays is a simple task. Standard
input power is 90-240Vac, 50/60Hz. DC input options include
9-18Vdc or 18-36Vdc.
DataChart
3600 Series Paperless Chart Recorders Features
DataChart
3600 Series Paperless Chart Recorders
Inputs & Computation Channels
Choose
6 or 12 universal, isolated direct inputs plus computation channels
for a total of 15.
- Measurement
types: linear, square root, log linear. T/C, RTD, frequency/pulse,
calculated, conditional, dry contact and external via serial
cornrnunications port.
- 3
scaleable voltage input ranges: +/- 100mV +/- 1 Volt and
+/- 10 Volts.
- Current
Ranges: 4-20mA and 10 - 50mA with internally switched 50
ohm shunts.
- Thermocouples:
B, C, E, J, K, R, S, T, Nickel/Nickel Moly and Nicrosil-Nisil.
- RTD's:
10 ohm Cu, 100 ohm Pt 385. 100 ohm Pt 392, 200 ohm Pt 385,
200 ohm Pt 392 and 120 ohm Nickel.
- Easy
setup and versatile scaling.
DataChart
3600 Series Paperless Chart Recorders Display
5
inch bright CCFL backlit LCD display with rugged Touch Screen
makes set up and operation simple. Browse data file information
from RAM or disk while recording. Real Time or historical data
review. Display modes include:
- Graphic
trend - vertical or horizontal
- Bargraphs
- vertical or horizontal
- Digital
meters
- Alarm/Event
Summary
- Split
screen combinations
Status
windows include:
- Time/Date
- Alarms
- Disk
capacity used
- Menu
DataChart
3600 Series Paperless Chart Recorders
High Accuracy Measurement
Isolated
input modules coupled with a fast 16 bit A/D converter provides
voltage measurement accuracy of +0.05%
DataChart
3600 Series Paperless Chart Recorders
Measurement Recording and Display Rates
- Each
channel is measured four times per second (250mS).
- Recording
rates are independently selectable in 1 second intervals
for each channel and range between 4 per second to 1 every
10 minutes
- Data
display update rates are selectable from 0.25 to 60 seconds.
- Virtual
chart speed is selectable from 0.5in to 600in/hr (10mm to
15000mm/hr).
DataChart
3600 Series Paperless Chart Recorders
High Quality Modular Design
- The
modular design of the DC-3600 allows options to be added
in the field quickly and easily.
- DataChart
Paperless Recorders and Accessories have a full two year
comprehensive warranty.
DataChart
3600 Series Paperless Chart Recorders
Programming Features
·
Menu selectable via intuitive Touch Screen
· Selectable signal input type and range.
· Application dependent scale ranging (up to + 5 digits).
· Selectable signal filtering.
· Engineering units - up to 5 character alphanumeric.
· Point tag - up to 10 characters (6 displayed).
· Program for normal or alarm/event triggered recording mode.
· User selectable multilevel passcode protection.
DataChart
3600 Series Paperless Chart Recorders
Computation Channels
- Math
functions include +, -, x, +, square root, totalization,
moving average, timed average, delta, gated timer, peak.
valley, logarithms, algebraic equations, conditional statements
and 12 programmable constants.
- Minimum
of 6 computation channels available.
- Computation
channels can be displayed, recorded and alarmed.
DataChart
3600 Series Paperless Chart Recorders
Data Storage
- 3,5inch
floppy disk (1.44Mbyte capacity, IBM format)
- PCMCIA
memory card (Flash to 220 Meg).
- Internal
128K RAM buffer provides historical data review on site
while still recording real time information. optional expansion
to 512K RAM is available.
- Search
data files by date/time or signal value.
- Transfer
data files to IBM compatible PC's via serial communications
port.
- Independent
data files for up to 12 channels, alarm/event file and configuration
files.
- Customize
file names for each recording session.
- Review
historical information while recording in real time
- Fill
to end mode with user selectable disk full alarm.
- Cyclical
storage mode.
DataChart
3600 Series Paperless Chart Recorders
Recorder Control
The
following functions can be controlled via external inputs or
internal alarms
| Function |
External |
Internal |
| Record
on/off |
Yes |
Yes |
| Record
rate |
Yes |
Yes |
| Chart
Speed |
Yes |
Yes |
| Totalizer
Reset |
Yes |
Yes |
| Event
message |
Yes |
No |
| Alarm
Ack/Reset |
Yes |
No |
| Scale
Set |
Yes |
No |
DataChart
3600 Series Paperless Chart Recorders Alarms
- 5
alarm set points per channel (60 total) include: high, low,
rate of change. true, false and abnormal signal conditions.
- 3
or 6 alarm contact outputs.
- Latching,
non-latching, failsafe, reflash and acknowledge.
DataChart
3600 Series Paperless Chart Recorders Communications
· RS232 bi-directional with modem support.
· RS4SS network port
- address up to 31 units
· Serial protocol
· MODBUS RTU
or MODBUS ASCII.
· Real time
and data file transfer to IBM compatible PC's.
· Remote configuration.
· User selectable baud rates from 300 to
19,200bps.