Calibration,Capacitor,Endevco,Model,2947B-2

Calibration Capacitor
Endevco Model 2947B-2

  Home | About Us | Products | Manufacturers | News | Contact

  Our Products

 Instrumentation
 Monitoring/Display
 Signal Conditioning


 Stress Analysis
 Strain Gauges


 Transducers
 Load/Torque Cells
 Displacement
 Pressure

 Data Acquisition

 Vibration
 Accelerometers
 Analysis


 Physical Testing
 Material/Composite
 Fatigue/Durability
 Physical


 Acoustics

 Environmental Test
 Chambers

 



      
  Our Manufacturers
 Ashcroft
 Capacitec
 Celesco
 Cincinnati Sub_Zero
 Data Physics
 Data Track
 Doric
 Ectron
 Endevco
 GP:50
 Honeywell
 Kaman
 Lebow Products
 Michigan Scientific
 Morehouse
 MTS Systems
 Non-Linear Devices
 Revere Transducers
 Sensortronics
 SRI-PMD
 Strainsert
 Tedea-Huntleigh
 Tovey
 Trans-Tek
 Unipulse
 Vishay Measurements
 Group
 
Web Site
www.intertechnology.com
 
  Calibration Capacitor Endevco Model 2947B-2
The Endevco Model 2947B-2 is a stable, precision, 1000 pF shielded capacitor designed specifically for calibrating charge sensing amplifiers. The voltage output from a signal generator is readily converted to a charge signal by connecting a capacitor in series between the generator and charge amplifier.

The transfer function is: Q = EC

Where:

Q = Charge, in picocoulombs (pC)

E = Voltage, in volts (V)

C = Capacitance, in picofarads (pF)

Since the capacitor is exactly 1000 pF, the transfer function simplifies to Q (pC) = E (mV). The charge must be expressed in units of pC pk if the voltage is expressed in peak units. Similarly, voltage rms converts to pC rms.

Features:

• Allows simulation of a piezoelectric transducer charge output
• Precision 1000 pF shielded capacitor
• Used for charge amplifier calibration
• Converts voltage to charge
• Easy to use


Endevco Model 2947B-2

  Home | About Us | Products | Manufacturers | News | Contact

Calibration Capacitor
Endevco Model 2947B-2


Site Designed And Maintained By Panther Marketing And Design